Possible Source
|
Test or Action
|
|
Improper dilution of highest standard, standard curve, or blank values*
|
Ensure use of appropriate diluent as blank value. Ensure accurate completion of 2-fold dilution series* **
|
|
Incomplete washing of wells
|
Ensure that wash apparatus is working correctly
|
|
Inadequate aspiration of wells
|
Wells should appear dry after aspiration
|
|
Unequal volumes added to wells
|
Check pipette function; recalibrate if necessary
|
|
Color Reagents A and B mixed too early*
|
Substrate Solution must be used within 15 minutes of mixing*
|
|
Pipetting error
|
Repeat assay
|